TITLE 32: ENERGY
CHAPTER II: ILLINOIS EMERGENCY MANAGEMENT AGENCY
SUBCHAPTER b: RADIATION PROTECTION
PART 335 MEDICAL USE OF RADIOACTIVE MATERIAL
SECTION 335.7090 THERAPY-RELATED COMPUTER SYSTEMS FOR MANUAL BRACHYTHERAPY


 

Section 335.7090  Therapy-related Computer Systems for Manual Brachytherapy  

 

The licensee shall:

 

a)         Perform acceptance testing on the treatment planning system of therapy-related computer systems in accordance with published protocols accepted by nationally recognized bodies.  At a minimum, the acceptance testing must include, as applicable, verification of:

 

1)         The source-specific input parameters required by the dose calculation algorithm;

 

2)         The accuracy of dose, dwell time, and treatment time calculations at representative points;

 

3)         The accuracy of isodose plots and graphic displays; and

 

4)         The accuracy of the software used to determine sealed source positions from radiographic images.

 

b)         Maintain a record of acceptance testing and copies of the protocols used for acceptance testing in accordance with this Section for 5 years after discontinuation of use of the treatment planning system.

 

(Source:  Added at 30 Ill. Reg. 9029, effective April 28, 2006)